FEATURES

  • Large LCD screen
  • Can display and save 50 groups of data, 5 groups per screen.
  • Data can be restored after a sudden power cut.
  • Auto setting Wavelength
  • Tungsten Halogen lamp & deuterium lamp can be turned on/off individually to extend lifetime.
  • Automatic WL. Calibration and dark current getting. With Scanning


Technical Specification

 

 

Optional System

Single - Beam, Grating

Wavelength Range

340 - 1000nm

Bandwidth

5nm

Wavelength Accuracy

±1nm

Wavelength Repeatability

0.5nm

Wavelength Setting

Auto

Photometric Accuracy

±0.5%T

Photometric Repeatability

0.3%T

Photometric Range

-0.3-3A,0-200%T

Stray Light

<0.3%T

Stability

±0.002A/h @500nm

Display

LCD

Detector

Silicon Photodiode

Standard Cell Holder

4-position 10 - 100 mm cell changer

Light Source

Tungsten Halogen Lamp & Deuterium lamp

Output

USB port & Parallel port (printer)

Power Requirements

AC 85 ~ 250V

Dimension

420x280x180mm

Weight

20Kg


Standard Configuration

*Main Set

:

1 No.

*USB Cable

:

1 No.

*Accessory kit. including a 4 glass cell (10 mm)

:

1 No.

*2 Quartz Cell (10 mm)

:

1 No.

*Operating Manual Software Manual

:

1 No.

*Dust Cover

:

1 No.

*Software CD

:

1 No.

*Power Cable

:

1 No.